Overview
Is medium exchange in a CPC (Cell Processing Center) a burden?
This cell culture system enables medium exchange within the bag while maintaining a closed system. Up to three units can be operated at the same time.
Sensors measure the liquid level in the bag, allowing precise control of medium input and output.


Features
1. Adopts a unique liquid thickness control method suitable for the bag culture
The pressure unit is equipped with a sensor to measure the thickness of the liquid in the bag, and while measuring the thickness of the liquid, it supplies new liquid and controls waste liquid. This method allows precise control of the liquid inside the bag despite its simple configuration.



2. Culture unit can be used in an incubator
A culture unit placed inside the incubator is connected via a cable to a control unit outs outside the incubator.

3. Microscopic Observation is possible while maintaining the closed system
The culture unit can be carried by removing the cable. Microscopic observation is possible through the glass surface of the pressure unit set on the culture unit.

4. Control by dedicated software
The culture system is controlled by dedicated software installed on the control computer. It is possible to set automatic culture medium exchange at any date and time.



5. Use a dedicated culture bag for the system
By using a WELLBAG dedicated to the system, it can be applied to complete medium exchange and differentiation induction during spheroid formation. In addition, the system-specific bag series, including a bag for adherent cells which currently under development, can be used.

Related Products
WELLBAG-S500-2PORTS
Medium Bag-S
(also used as a waste bag)
Liquid Transfer Tube-S
Product Numbers
Item
| Item | Product Code | Quantity | |
|---|---|---|---|
| CELL CULTURE SYSTEM ※ACS1000SET |
![]() |
71000-01 | ①Control unit ×1 (ACS1000) →Size:W360 × D300 × H88 mm ②Culture unit ×3 (CU1000) →Size:W280 × D173 × H60 mm ③Holder unit ×3 (HU1000) ④Control PC ×1 (Special software installed) |
Consumables
| Consumables | Product Code | Quantity | |
|---|---|---|---|
| WELLBAG-S500-2PORTS ※For system use, Consumable →Culture area:50cm² →Culture volume:20~20ml |
![]() |
11010-05 | Pack of 5 bags |
| 11010-10 | Pack of 10 bags | ||
| 11010-00 | Pack of 100 bags | ||
| Medium bag-S (Dual-use for Supply/Waste) ※Consumables |
![]() |
B1000-05 | 5 sets / pack (For 10 bags) |
| B1000-10 | 10 sets / pack (For 20 bags) | ||
| B1000-00 | 100 sets / pack (For 200 bags) | ||
| Pumping Tube-S ※For peristaltic pump ※Consumables |
![]() |
T1000-05 | 5 sets / pack (For 10 tubes) |
| T1000-10 | 10 sets / pack (For 20 tubes) | ||
| T1000-00 | 100 sets / pack (For 200 tubes) | ||
Product Documents
Application
Coming soon!
Videos
CellCultureSystem Overview Video
FAQ
Please contact us.
We have confirmed that it can be performed for up to 40 days.
No. We recommend wiping it down with alcohol, hydrogen peroxide, or similar agents.
Please refer to the evaluation example below.
Media exchange performance in a well bag (for system use): Example
By using 14.25 mL of medium for a 10 mL well bag, a 99% media exchange was achieved.
The migration rate of spheroids within the well bag during exchange was 0.5% or less.

Test conditions
Flow rate: 0.2 mL/min
Medium volume in the media bag: 10 mL